| Low-pressure HVAC | 0 to 2.5 kPa (0 to 10 inH₂O) | ≤ ±0.05% of reading | 0.1 Pa to 1 Pa | High sensitivity, stable zero, low-temperature drift, and compatibility with differential or gauge pressure. | Low-pressure differential calibrator with fine adjustment and a stable internal pressure sensor. |
| Cleanroom monitoring | 0 to 1 kPa (0 to 4 inH₂O) | ≤ ±0.25% of full scale | 1 Pa or better | Repeatability, low leakage, data logging, and the ability to verify small pressure differences. | Portable differential calibrator with a low-pressure range, digital display, and documented stability. |
| Pneumatic control systems | 0 to 100 kPa (0 to 1 bar / 15 psi) | ≤ ±0.05% of full scale | 10 Pa to 100 Pa | Gauge pressure measurement, fast response, reliable pressure generation, and compatibility with common fittings. | Hand-operated or motor-assisted pressure calibrator with adjustable output and leak-check capability. |
| Process instruments | 0 to 1 MPa (0 to 10 bar / 145 psi) | ≤ ±0.05% of reading | 100 Pa to 1 kPa | Measurement uncertainty should be lower than the tolerance of the device under test; temperature compensation is beneficial. | Calibrator with a suitable pressure module, fine pressure control, stable output, and calibration traceability. |
| Industrial pneumatic lines | 0 to 1.6 MPa (0 to 16 bar / 232 psi) | ≤ ±0.1% of full scale | 1 kPa or better | Rugged construction, overpressure protection, quick connections, and resistance to field contamination. | Field pressure calibrator with reinforced ports, replaceable seals, and a clear pressure readout. |
| Pressure switches | Vacuum to 1 MPa (approximately −1 to 10 bar) | ≤ ±0.05% of reading | 100 Pa to 1 kPa | Adjustable ramping, switch-state detection, pressure stabilization, and programmable test points. | Pressure switch tester with automated step generation, electrical input/output monitoring, and test reporting. |
| Safety and alarm testing | 0 to 2.5 MPa (0 to 25 bar / 363 psi) | ≤ ±0.05% of full scale | 1 kPa or better | Controlled pressure increase, repeatable set-point testing, pressure relief, and secure connections. | Calibrator with a high-pressure pneumatic pump, fine volume control, and suitable overpressure safeguards. |
| Vacuum instruments | 0 to −100 kPa (0 to −1 bar / 0 to −14.5 psi) | ≤ ±0.1% of full scale | 100 Pa to 1 kPa | Vacuum stability, leak-rate control, atmospheric reference compensation, and appropriate vacuum-rated seals. | Vacuum calibrator with a controlled vacuum source, fine adjustment valve, and gauge or absolute reference mode. |
| Absolute pressure measurement | 80 to 120 kPa absolute (typical atmospheric region) | ≤ ±0.05% of reading | 10 Pa to 100 Pa | Absolute reference, barometric compensation, stable temperature performance, and adequate atmospheric-pressure accuracy. | Absolute pressure calibrator with an integrated or externally verified barometric reference. |
| Laboratory calibration | Selected according to the device under test | Typically 3:1 or better accuracy ratio | At least 10 times finer than the test tolerance | Traceability, repeatability, hysteresis evaluation, temperature control, and complete uncertainty budgeting. | High-resolution reference calibrator with documented calibration coefficients, stable pressure generation, and digital records. |